Replacement Oxygen Sensor for OXYPro® WR / WRM

Oxygen Exchange Cap OEC-PSt7-YAU

The Oxygen Exchange Cap is made of stainless steel and has an integrated sensor type PSt7. The sensor has a maximum measurement range of 0 - 100 % O2, and a detection limit of 0.03 % O2. It stands steam sterilization and cleaning in place. The OEC-PSt7-YAU is available in different versions: USP Class VI compatible (YOP) for bioprocess monitoring, safe for food applications (CAF), or with a hydrophobic Teflon coating (PTFE) for measurements in aggressive environments. The OEC can be screwed to the top of the OXYPro® WR / WRM fitting. Exchange of a used sensor is made easy as sensor calibration can be done via one fast barcode scan.

  • Maximum measurement range of 0 - 100 % O2
  • Different versions: USP Class VI compatible, food applications, aggressive environments
  • Stands SIP (max. + 140 °C, 1.5 atm) & CIP (2 % NaOH, + 80 °C)
  • Easy exchange due to barcode calibration
  • Pressure resistant
Get a quote Place an order

Technical

Specifications OEC-PSt7-YAU-YOP OEC-PSt7-YAU-CAF OEC-PSt7-YAU-PTFE
*at 20 °C, 960 - 980 hPa; humidified gas mixture
**after customized calibration
***determined changing from 20.9 % oxygen gas to nitrogen 5.0
****determined changing from air-saturated water to a freshly prepared 1 % sodium sulfite solution


 
Measurement range* Optimal: 0 - 50 % O2, 0 - 500 hPa, 0 - 22.5 mg/L
Max.**: 0 - 100 % O2, 0 - 1000 hPa, 0 - 45.0 mg/L
Limit of detection* 0.03 % O2, 0.3 hPa, 0.015 mg/L
Resolution*

1 % O2 ± 0.02 % O2, 20.9 % O2 ± 0.1 % O2
0.4 mg/L ± 0.009 mg/L, 9 mg/L ± 0.04 mg/L

Response time (t90)
Gaseous O2***
Dissolved O2****

< 10 sec.
< 30 sec.

< 15 sec.
< 40 sec.

< 85 sec.
< 100 sec.
Properties      
Compatibility Aqueous solutions
No cross-sensitivity to pH 1 - 14, CO2, SO2, ionic species
Cross-sensitivity to Chlorine gas
Organic solvents such as pure acetone, toluene, chloroform or methylene chloride
Steam sterilization (max. + 140 °C, 1.5 atm) Yes Yes Yes
Cleaning in place (CIP) (+ 80 °C, 2 % NaOH) Yes Yes Yes
Cleaning procedure 3 % H2O2, acidic agents (HCl, H2SO4) max. 4 - 5 %
Ethanol, methanol, cleaning agents at room temperature
Calibration Two-point calibration in oxygen-free (e. g. nitrogen 5.0, 1 % sodium sulfite solution) and air-saturated environment (20.9 % oxygen, air-saturated water)
Storage stability 5 years provided the sensor material is stored at room temperature in dry conditions and in the dark
Pressure resistance* - 1 bar to + 12 bar
Temperature range operation Optimal: 0 °C to + 50  °C
Max.: - 10 °C to + 90 °C
Storage conditions

Optimal: room temperature (+ 20 °C ± 5 °C)
Max.: 0 °C to  + 70 °C, 80 % rel. humidity

Variants USP Class VI compatible Safe for food applications Hydrophobic Teflon coating

Related products

Wide Range Oxygen Probe OXYPro® WR (max. 0 - 100 % O2)

Wide Range Oxygen Probe OXYPro® WR, WR-ECS & WRM
(max. 0 - 100 % O2)

The OXYPro® WR & WRM combine an optical sensor type PSt7 and…