Top

Reduces Air Bubble Formation during Dissolved Oxygen Measurements

Oxygen Exchange Cap OEC30-PSt3-YAU

This beveled sensor cap with integrated PSt3 sensor reduces air bubble formation on the sensor during dissolved oxygen measurements with an oxygen probe for in-line measurement (OIM). The PSt3 sensor has a maximum measurement range of 0 - 100 % O2, and a detection limit of 0.03 % O2. The sensor cap is available in different versions: USP Class VI compatible (YOP) for bioprocess monitoring, safe for food applications (CAF), or with a hydrophobic Teflon coating (PTFE) for measurements in aggressive environments. It stands steam sterilization and cleaning in place.

  • Maximum measurement range of 0 - 100 % O2
  • Reduced air bubble formation
  • Different versions: USP Class VI compatible, food applications, aggressive environments
  • Stands SIP (max. + 140 °C, 1.5 atm) & CIP (+ 80 °C, 2 % NaOH)
  • Easy exchange due to barcode calibration
  • Pressure resistant
Get a quote Place an order

Technical

SpecificationsOEC30-PSt3-YAU-YOPOEC30-PSt3-YAU-CAFOEC30-PSt3-YAU-PTFE
*at 20 °C, 960 - 980 hPa; humidified gas mixture
**after customized calibration
***after two-point calibration as described in the manual
****determined changing from 20.9 % oxygen gas to nitrogen 5.0
*****determined changing from air-saturated water to a freshly prepared 1 % sodium sulfite solution
Measurement range*Optimal: 0 - 50 % O2, 0 - 500 hPa, 0 - 22.5 mg/L
Max.**: 0 - 100 % O2, 0 - 1000 hPa, 0 - 45 mg/L
Limit of detection*0.03 % O2, 0.3 hPa, 0.020 mg/L
Resolution*1 % O2 ± 0.02 % O2, 20.9 % O2 ± 0.1 % O2
0.4 mg/L ± 0.009 mg/L, 9 mg/L ± 0.04 mg/L
Accuracy at 20 °C***1 ± 0.05 % O2, 20.9 ± 0.2 % O2
Response time (t90)
Gaseous O2****
Dissolved O2*****

< 10 sec.
< 30 sec.

< 15 sec.
< 40 sec.

< 85 sec.
< 100 sec.
Properties   
CompatibilityAqueous solutions, ethanol, methanol
No cross-sensitivitypH 1 - 14, CO2, H2S, SO2, ionic species
Cross-sensitivityChlorine gas
Organic solvents such as pure acetone, toluene, chloroform or methylene chloride
Steam sterilization (+ 140 °C, 1.5 atm)YesYesYes
Cleaning in place (CIP)
(+ 80 °C, 2 % NaOH)
YesYesYes
Cleaning procedure3 % H2O2, acidic agents (HCl, H2SO4) max. 4 - 5 %
Cleaning agents containing ClO2 at room temperature
CalibrationTwo-point calibration in oxygen-free (e. g. nitrogen 5.0, 1 % sodium sulfite solution) and air-saturated environment (20.9 % oxygen, air-saturated water)
Storage stability5 years provided the sensor material is stored at room temperature in dry conditions and in the dark
Pressure resistance*10 mbar up to 12 bar
Temperature range operationOptimal: 0 °C to + 50 °C
Max.: - 40 °C to + 90 °C
Temperature range storageOptimal: room temperature (+ 20 °C ± 5  °C)
Max.: 0 °C to + 70 °C
VariantsUSP Class VI compatibleSafe for food applicationsHydrophobic Teflon coating

Related products

OIM-PSt3 Oxygen Probe for In-Line Measurement

Oxygen Probe for In-Line Measurement OIM-PSt3

The oxygen probe for in-line measurement (OIM) consists of a…

More

Fibox 4

The battery powered Fibox 4 is designed for easy handheld…

More
Fibox 4 trace handheld fiber optic O2 meter

Fibox 4 trace

The Fibox 4 trace is a stand-alone oxygen meter designed for…

More
Wall mount optical oxygen meter OXY-1 WM

OXY-1 WM

The OXY-1 WM is a precise single-channel oxygen meter for…

More
Wall mount optical trace oxygen meter OXY-1 WM trace

OXY-1 WM trace

The OXY-1 WM trace is a precise single-channel oxygen meter…

More
OXY-4 SMA 4-channel optical O2 meter

OXY-4 SMA (G3)

This multi-channel oxygen meter is ideally suited for…

More

OXY-4 SMA trace (G3)

The OXY-4 SMA trace is compatible with sensor spots, dipping…

More

Presens TV

Watch tutorials, webinars and informative videos about PreSens optical sensor systems.

All videos