Probe for Dissolved & Gaseous Oxygen Measurements

OXYPro-PSt7

The OXYPro is connected directly to the controller. It combines an optical sensor and electro-optical module in one probe. OXYPro has a standard PG 13.5 thread and stainless steel housing. As a digital interface it uses RS485 (PreSens proprietary or Modbus RTU) and has 4 - 20 mA analog output. A serial communication protocol is offered for data exchange between OXYPro and a PC or another host unit. For initial operation and adjustment of the 4 - 20 mA outputs the PreSens EOM-STS software can be used.The optical oxygen sensor type PSt7 has a maximum measurement range of 0 - 100 % O2, and a detection limit of 0.03 % O2. The removable sensor cap (Optical Exchange Cap OEC) can easily be exchanged. The probe is pre-calibrated, stands steam sterilization (SIP) and cleaning in place (CIP).

  • Maximum measurement range of 0 - 100 % O2, detection limit 0.03 % O2
  • No polarization time
  • Stands SIP (max. 140 °C, 1.5 atm) & CIP (2 % NaOH, + 80 °C)
  • Simple integration due to standard PG 13.5 thread
  • Communication via RS485 (PreSens proprietary or Modbus RTU)
  • 4 - 20 mA analog output

Applications

OXYPro for Your Bioreactor

This optical oxygen probe can easily be integrated in your control systems. The precise oxygen measurements enable control and contribute to safe and stable culture conditions. The OXYPro can be mounted to standard ports, and stands cleaning in place and steam sterilization. No membrane cleaning, polarization or replenishment of electrolyte solutions is necessary. A used sensor cap can easily be exchanged. This low maintenance probe reduces the work load considerably.


Technical

Specifications
Gaseous & Dissolved O2
Dissolved O2
*at 20 °C, 960 - 980 hPa; humidified test gas mixture
**after customized calibration
***determined changing from 20.9 % oxygen gas to nitrogen 5.0
****determined changing from air saturated water to a freshly prepared 1 % sodium sulfite solution
Oxygen sensorPSt7
Measurement range*

Optimal: 0 - 50 % O2
       0 - 500 hPa
Max.**:   0 - 100 % O2
         0 - 1000 hPa

Optimal: 0 - 25 mg/L
       0 - 780 µmol/L
Max.**:   0 - 45 mg/L
         0 - 1400 µmol/L

Limit of detection*0.03 % O220 ppb
Resolution*1 ± 0.02 % O2
20.9 ± 0.1 % O2
50 ± 0.3 % O2
Accuracy at + 20 °C1 ± 0.05 % O2, 20.9 ± 0.2 % O2
Response time (t90)< 10 sec.***< 30 sec.****
Maximum pressure12 bar
Properties
CompatibilityAqueous solutions, ethanol, methanol, cleaning agents containing ClO2 at room temperature
Cross-sensitivityOrganic solvents, such as acetone, toluene, chloroform or methylene chloride
Chlorine gas
Sterilization procedureSteam sterilization (max. 140 °C, 1.5 atm)
Cleaning procedureCleaning in place (CIP, 2 % NaOH, + 80 °C, + 176 °F)
CalibrationTwo-point calibration with oxygen-free environment (nitrogen 5.0, 1 % sodium sulfite) and air-saturated environment (20.9 % oxygen, or air-saturated water)
Storage stability of sensor cap5 years, provided the sensor is stored at room temperature in dry conditions and in the dark
Sampling ratefrom 1 sec. up to 9 min. 59 sec.
Temperature sensorNTC
Temperature performanceAccuracy below ± 1 °C
Power supply7 - 30 V

Power consumption in active modemax. 1 W
Power consumption in stand-by mode0.15 W
Temperature range operationoptimal from 0 °C to + 50 °C / maximal from - 40 °C to + 90 °C**
Temperature range storagefrom - 10 °C to + 70 °C
Dimensions12 mm x 120 mm with PG 13.5 thread and VP8 connector
Surface quality of steelRa < 0.4 µm (N5)
Wetted partstainless steel 1.4435, silicone, EPDM
Weight100 g
Digital interfaceRS485, half duplex (variable Baud rate, default: 19200, data bits: 8, parity: none, stop bits: 1, handshake: none)
Optional: RS485 Modbus RTU
Analog output4 - 20 mA