Ultra-low oxygen measurements with metal flow-through cells
Oxygen Exchange Cap OEC-M10-PSt9
This oxygen exchange cap holds a PSt9 oxygen sensor (measurement range 0 - 200 ppmv O2, limit of detection 0.5 ppmv O2) for measurements of ultra-low oxygen traces in gas pipes. The OEC-M10 can be screwed to a FTM-M10 via its M10 thread. This way a used sensor coating in your flow-through cell can easily be replaced. A polymer optical fiber with protective metal spiral tube cover can be attached to this OEC. The PSt9 sensor is only suitable for measurements in the gas phase.
- Replacement sensor for FTM-M10
- Detect ultra-low oxygen traces in gas pipelines
- Max. measurement range 0 - 200 ppmv O2, limit of detection 0.5 ppmv O2
- Signal independent from flow velocity
Technical
Specifications | Gaseous O2 | |||
---|---|---|---|---|
*at 20 °C, 960 - 980 hPa; humidified gas mixture **determined changing from 20.9 % oxygen gas to nitrogen 5.0 | ||||
Measurement range* | 0 - 200 ppmv O2 | |||
Limit of detection* | 0.5 ppmv O2 | |||
Resolution* | 10 ± 0.5 ppmv O2 100 ± 0.8 ppmv O2 200 ± 1.5 ppmv O2 | |||
Response time (t90)** | < 10 sec. | |||
Properties | ||||
Compatibility | Gas phase only! | |||
No cross-sensitivity | CO2, H2S, SO2 | |||
Cross-sensitivity | Organic vapor Chlorine gas | |||
Calibration | Two-point calibration with oxygen-free environment (nitrogen 6.0) and a second calibration value optimally between 100 and 200 ppmv gaseous oxygen | |||
Temperature range operation | Optimal: 0 °C to + 40 °C Max.: - 20 °C to + 60 °C | |||
Temperature range storage | Optimal: room temperature (+ 20 °C ± 5 °C) Max.: 0 °C to + 70 °C | |||
Storage stability | 5 years provided the sensor is stored at room temperature in dry conditions and in the dark |