Top

Mid-Range Oxygen Sensor with Fast Response Time

Oxygen Exchange Cap OEC-PSt8-NAU

The Oxygen Exchange Cap with oxygen sensor type PSt8 is made of stainless steel. It has a maximum measurement range of 0 - 20.9 % O2, and a detection limit of 0.007 % O2. This sensor cap is specifically designed for applications where a fast sensor response time (< 3 sec.) is needed. It is screwed to the top of the OXYPro® MR / MRM fitting to replace a used sensor. It can easily be calibrated with one fast barcode scan. OEC-PSt8-NAU-OIW is not autoclavable.

  • Maximum measurement range of 0 - 20.9 % O2
  • Fast response time (< 3 sec.)
  • Easy exchange due to barcode calibration
  • Pressure resistant
Get a quote Place an order

Technical

Specifications OEC-PSt8-NAU-OIW
*at 20 °C, 960 - 980 hPa; humidified gas mixture
**after customized calibration
***determined changing from 10 % oxygen gas to nitrogen 5.0
****determined changing from air-saturated water to a freshly prepared 1 % sodium sulfite solution
Measurement range* Optimal: 0 - 10 % O2, 0 - 100 hPa, 0 - 4.5 mg/L
Max.**: 0 - 20.9 % O2, 0 - 200 hPa, 0 - 9.0 mg/L
Limit of detection* 0.007 % O2, 0.07 hPa, 3 ppb
Resolution* 0.007 % O2 ± 0.002 % O2, 0.023 % O2 ± 0.005 % O2, 2.314 % O2 ± 0.007 % O2
3 ppb ± 1 ppb, 10 ppb ± 2 ppb, 1000 ppb ± 3 ppb
Response time (t90)
Gaseous O2***
Dissolved O2****

< 3 sec.
< 3 sec.
Properties  
Compatibility Aqueous solutions
No cross-sensitivity to pH 1 - 14, CO2, SO2, ionic species
Cross-sensitivity to Chlorine gas
Organic solvents such as pure acetone, toluene, chloroform or methylene chloride
Cleaning procedure 3 % H2O2, acidic agents (HCl, H2SO4) max. 4 - 5 %
Ethanol, methanol, cleaning agents at room temperature
Calibration Two-point calibration in oxygen-free environment (e. g. nitrogen 5.0, 1 % sodium sulfite solution) and a second calibration medium between 2 and 4 % oxygen
Storage stability 5 years provided the sensor material is stored at room temperature in dry conditions and in the dark
Pressure resistance* - 1 bar to + 12 bar
Temperature range operation Optimal: 0 °C to + 50 °C
Max.: - 10 °C to  + 70 °C
Storage conditions Optimal: room temperature (+ 20 °C ± 5 °C)
Max.: 0 °C to + 70 °C, 80 % rel. humidity

Related products

Mid-Range Oxygen Probe OXYPro® MR (max. 0 - 20.9 % O2)

Mid-Range Oxygen Probe OXYPro® MR & MRM
(max. 0 - 20.9 % O2)