Ultra-Trace Oxygen Sensor Replacement for OIM and OXYPro® UT / UTM
Oxygen Exchange Cap OEC-PSt9-NAU
A PSt9 sensor is integrated in a stainless steel cap, which can be used to replace a used sensor on an OIM (Oxygen Probe for In-line Measurement) or the OXYPro® UT / UTM. The ultra-trace sensor has a measurement range of 0 - 200 ppmv oxygen with a detection limit of 0.5 ppmv, so even smallest gaseous oxygen traces can be detected. The sensor is not autoclavable. It is ideally suited for monitoring oxygen sensitive applications or production processes in different industries.
- Measurements down to 0.5 ppmv oxygen (gaseous O2 only!)
- Fast response time
- Easy exchange due to barcode calibration
- Pressure resistant
Technical
Specifications | OEC-PSt9-NAU |
---|---|
*at 20 °C, 960 - 980 hPa; humidified gas mixture **determined changing from 200 ppmv oxygen gas to nitrogen 6.0 | |
Measurement range* | 0 - 200 ppmv O2 |
Limit of detection* | 0.5 ppmv O2 |
Resolution* | 1 ± 0.15 ppmv O2, 100 ± 0.8 ppmv O2, 200 ± 1.5 ppmv O2 |
Response time (t90)** | < 10 sec. |
Properties | |
Compatibility | Gas phase only! |
No cross-sensitivity to | CO2, SO2 |
Cross-sensitivity to | Chlorine gas Organic solvents such as pure acetone, toluene, chloroform or methylene chloride |
Calibration | Two-point calibration in oxygen-free environment (e. g. nitrogen 5.0) and a second calibration medium between 100 and 200 ppmv oxygen |
Storage stability | 5 years provided the sensor is stored at room temperature in dry conditions and in the dark |
Pressure resistance* | 3 bar |
Temperature range operation | 0 °C to + 40 °C |
Storage conditions | 0 °C to + 70 °C, 80 % rel. humidity |
Variant | USP Class VI compatible |