Watch tutorials, webinars and informative videos about PreSens optical sensor systems.
Steel Probe for Monitoring Ultra-Low Oxygen Traces in Gases
Oxygen Probe for In-Line Measurements OIM-PSt9
The OIM-PSt9 has a stainless steel housing with standardized thread, so it is compatible with most ports and port adapters. The optical ultra-trace oxygen sensor type PSt9 has a measurement range of 0 - 200 ppmv oxygen, and a detection limit of 0.5 ppmv O2. It is integrated in a removable sensor cap, which is screwed to the top of the metal housing and can be replaced if needed. The probe can be equipped with customized fittings for e. g. sight glasses. OIM-PSt9 is ideally suited for gaseous O2 monitoring in oxygen sensitive processes, and together with the OXY-1 WM trace it can also be applied in industrial environments.
- Measurement range of 0 - 200 ppmv O2, detection limit 0.5 ppmv O2
- For gaseous O2 measurements only
- Pressure resistant
- Ultra-trace O2 monitoring in gases, sensitive processes
Technical
Specifications | Gaseous O2 |
---|---|
Measurement range | 0 - 200 ppmv oxygen |
Limit of detection | 0.5 ppmv oxygen |
Resolution | 10 ± 0.5 ppmv oxygen 100 ± 0.8 ppmv oxygen 200 ± 1.5 ppmv oxygen |
Accuracy at + 20 °C* | ± 2 ppmv oxygen or ± 5 %, whichever is higher |
Measurement temperature range | from 0 to + 40 °C |
Response time (t90) | < 3 sec. |
Properties | |
Compatibility | Gas phase only |
No cross-sensitivity | CO2, SO2 |
Cross-sensitivity | Organic vapor, Chlorine gas |
Calibration | Two-point calibration with oxygen-free environment (nitrogen 6.0) and a second calibration value optimally between 100 and 200 ppmv gaseous oxygen |
Storage stability | 24 months provided the sensor material is stored in the dark at room temperature |
*after two-point calibration as described in the manual |