Oxygen Exchange Cap OEC30-PSt3-NAU
A wide range PSt3 sensor is integrated in a beveled stainless steel cap. This cap reduces air bubble formation on the sensor. It has a very fast response time (< 3 sec.) and a maximum measurement range of 0 - 100 % O2 (detection limit 0.03 % O2 / 20 ppb). The OEC30-PSt3-NAU is screwed to the OXYBase® WR housing to replace a used sensor and can be calibrated with one fast barcode scan. This sensor cap is not autoclavable.
- Maximum measurement range of 0 - 100 % O2
- Reduced air bubble formation
- Fast response time (< 3 sec.)
- Easy exchange due to barcode calibration
- Pressure resistant
Technical
Specifications | OEC30-PSt3-NAU-OIW |
---|---|
*at 20 °C, 960 - 980 hPa; humidified gas mixture **after customized calibration ***determined changing from 20.9 % oxygen gas to nitrogen 5.0 ****determined changing from air-saturated water to a freshly prepared 1 % sodium sulfite solution | |
Measurement range* | Optimal: 0 - 50 % O2, 0 - 500 hPa, 0 - 22.5 mg/L Max.**: 0 - 100 % O2, 0 - 1000 hPa, 0 - 45.0 mg/L |
Limit of detection* | 0.03 % O2, 0.3 hPa, 0.020 mg/L |
Resolution* | 1 % O2 ± 0.02 % O2, 20.9 % O2 ± 0.1 % O2 0.4 mg/L ± 0.009 mg/L, 9 mg/L ± 0.04 mg/L |
Response time (t90) Gaseous O2*** Dissolved O2**** | < 3 sec. < 3 sec. |
Properties | |
Compatibility | Aqueous solutions, ethanol, methanol |
No cross-sensitivity to | pH 1 - 14, CO2, H2S, SO2, ionic species |
Cross-sensitivity to | Chlorine gas Organic solvents such as pure acetone, toluene, chloroform or methylene chloride |
Cleaning procedure | 3 % H2O2, acidic agents (HCl, H2SO4) max. 4 - 5 % Cleaning agents containing ClO2 at room temperature |
Calibration | Two-point calibration in oxygen-free (e. g. nitrogen 5.0, 1 % sodium sulfite solution) and air-saturated environment (20.9 % oxygen, air-saturated water) |
Storage stability | 5 years provided the sensor material is stored at room temperature in dry conditions and in the dark |
Pressure resistance* | 3 bar |
Temperature range operation | Optimal: 0 °C to + 40 °C Max.: - 10 °C to + 90 °C |