Watch tutorials, webinars and informative videos about PreSens optical sensor systems.
Measure pH with High Spatial Resolution
Needle-type pH Microsensor NTH-HP5-ST-high
This pH microsensor is ideally suited for measurements at high pH range (6 – 9 pH). The NTH is based on a 250 μm plastic fiber which is very resistant against damage or breaking. With its protective syringe needle housing the NTH-HP5-ST-high can easily penetrate tissue, septum rubber or packaging materials. When the sensor is in place the sensor tip is extended via the syringe plunger for measurements.
- Robust plastic fiber
- For measurements at high pH (6 – 9 pH)
- High spatial resolution
- Measurement in smallest volumes
- Integration into plant and animal tissue
- Optimized for culture media and physiological solutions
Applications
pH Measurement in Plants & Animals
pH Microsensors can be implemented in plant tissue or even in small animals. New insights in physiological aspects can be obtained. Due to the small size of the probes only a minimal disturbance will occur.
pH Measurement in Small Volumes
Due to the small dimension of the probe, pH measurements can be done in very small volumes - even in microtiter plates of a higher format like 384 or 1536. No need for reference electrodes - a real step forward, especially in small volumes. Of course, the measurement is independent of electromagnetical fields - this even allows measuring in NMR spectrometers.
Technical
Specifications* | |
---|---|
*provided pH sensors are used without further handling in physiological solutions **equilibrated sensor kept in well stirred solution at + 37 °C | |
Measurement range | 6 - 9 pH |
Resolution | at pH = 7: ± 0.02 pH |
Drift | at pH = 7: < 0.1 pH per day (sampling interval of 1 min.) |
Measurement temperature range | from + 5 to + 50 °C |
Response time (t90)** | at + 37 °C: < 30 sec. |
Properties* | |
Compatibility | Aqueous solutions, ethanol (max. 10 % v/v), pH 2 - 10 |
No cross-sensitivity | Electrical fields, proteins |
Cross-sensitivity | Reduced to ionic strength (salinity); a high concentration of small fluorescent molecules in the visible range can interfere |
Disinfection procedure | One-time disinfection with EtOH (70 % v/v) or isopropanol (100 % v/v) |
Cleaning procedure | Buffer, water, pepsin solution |
Calibration | One-point or multi-point calibration procedure supported by PreSens Measurement Studio 2 software |
Related products
Resources
FAQs
Manuals
Brochures
Software
PreSens Measurement Studio 2 version 4.0.0 - Windows 10/11
PreSens Profiling Studio version 2.0.0 - Windows 10/11