Watch tutorials, webinars and informative videos about PreSens optical sensor systems.
Trace Oxygen Measurement in Metal Flow-Through Connectors FTM
Oxygen Exchange Window OEW-PSt6
The Oxygen Exchange Window is integrated in the FTM-PSt6. It enables to easily exchange a used sensor coating without having to replace the whole flow-through connector. The integrated PSt6 type sensor allows trace oxygen measurements in gases and liquids. It has a measurement range of 0 – 5 % oxygen, with a detection limit of 0.002 % oxygen.
- Easy exchange of a used sensor coating
- Online oxygen measurement in perfusion systems
- No consumption of oxygen
- Signal independent of flow velocity
- Autoclavable, SIP (+ 130 °C, 1.5 atm) & CIP (2 % NaOH, + 80 °C)
- Polarization-free
- Long-term stability
Applications
Process Control in Biotech & Pharma
The OIMs, the oxygen probes for in-line measurement, are designed for harsh measurement conditions. These systems give high accuracy and are almost maintenance-free as they do not contain any membranes and electrolytes. The sensors are fully autoclavable and sterilizable (CIP & SIP).
In-line Oxygen Measurement in Brewing and Beverage Industry
Even low concentrations of oxygen influence shelf-life and the taste of certain beverages. Due to the extraordinary low detection limit and outstanding accuracy of PreSens trace oxygen sensors, the quality of oxygen-sensitive products such as beer, wine and soft drinks is secured.
The optical sensors have – compared to standard electrodes – a very fast response time and they are very easy to maintain.
Technical
Specifications | Gaseous & Dissolved O2 | Dissolved O2 |
---|---|---|
*after two-point calibration as described in the manual | ||
Measurement range | 0 – 5 % O2 0 – 41.4 hPa | 0 – 2 mg/L 0 – 56.9 µmol/L |
Limit of detection | 0.002 % oxygen | 1 ppb |
Resolution | ± 0.0007 % O2 at 0.002 % O2 | ± 0.0003 mg/L at 0.001 mg/L |
Accuracy* | ± 1 ppb or ± 3 % of the respective concentration; whichever is higher | |
Drift at 0 % oxygen | < 2 ppb within 30 days (sampling interval of 1 min.) | |
Measurement temperature range | from 0 to + 50 °C | |
Response time (t90) | < 6 sec. | < 40 sec. |
Properties | ||
Compatibility | Aqueous solutions, ethanol, methanol | |
No cross-sensitivity | pH 1 – 14 | |
Cross-sensitivity | Organic solvents, such as acetone, toluene, chloroform or methylene chloride Chlorine gas | |
Sterilization procedure | Steam sterilization | |
Cleaning procedure | Cleaning in place (CIP, 2 % NaOH, + 80 °C, + 176 °F) | |
Calibration | Two-point calibration with oxygen-free environment (nitrogen, sodium sulfite) and a second calibration value optimally between 1 % and 2 % oxygen | |
Storage stability | 24 months provided the sensor material is stored in the dark |
Related products
Resources
FAQs
Manuals
Brochures
Software
PreSens Datamanager version 2.0.0.57 - Windows XP/Vista/7/8