Oxygen Probe Integrated in Metal Flow-Through Connector FTM-PSt3
An oxygen exchange window is integrated in a metal flow-through connector. The FTM can be integrated in pipes with o.d. 6 mm or other diameters using the respective adapters. The oxygen probe is very robust and long-term stable. It can be cleaned by CIP (Cleaning in place 2 % NaOH, + 80 °C) and also stands steam sterilization (+ 130 °C, 1.5 atm).
- Online oxygen measurement in perfusion systems
- No consumption of oxygen
- Signal independent of flow velocity
- Autoclavable, SIP (+ 130 °C, 1.5 atm) & CIP (2 % NaOH, + 80 °C)
- Polarization-free
- Long-term stability
Applications
In-line Oxygen Measurement in Brewing and Beverage Industry
Even low concentrations of oxygen influence shelf-life and the taste of certain beverages. Due to the extraordinary low detection limit and outstanding accuracy of PreSens trace oxygen sensors, the quality of oxygen-sensitive products such as beer, wine and soft drinks is secured.
The optical sensors have – compared to standard electrodes – a very fast response time and they are very easy to maintain.
Technical
Specifications | Gaseous & Dissolved O2 | Dissolved O2 |
---|---|---|
*after two-point calibration as described in the manual | ||
Measurement range | 0 – 100 % O2 0 – 1000 hPa | 0 – 45 mg/L 0 – 1400 µmol/L |
Limit of detection | 0.03 % oxygen | 15 ppb |
Resolution | ± 0.01 % O2 at 0.21 % O2 | ± 0.004 mg/L at 0.091 mg/L |
Accuracy* | ± 0.4 % O2 at 20.9 % O2 ± 0.05 % O2 at 0.2 % O2 | |
Drift | < 0.03 % O2 within 30 days (sampling interval of 1 min. / at 0% oxygen) | |
Measurement temperature range | from 0 to + 50 °C | |
Response time (t90) | < 6 sec. | < 40 sec. |
Properties | ||
Compatibility | Aqueous solutions, ethanol, methanol | |
No cross-sensitivity | pH 1 – 14 | |
Cross-sensitivity | Organic solvents, such as acetone, toluene, chloroform or methylene chloride Chlorine gas | |
Sterilization procedure | Steam sterilization | |
Cleaning procedure | Cleaning in place (CIP, 2 % NaOH, + 80 °C, + 176 °F) 3 % H2O2 Acidic agents (HCl, H2SO4), max. 4 – 5 % | |
Calibration | Two-point calibration with oxygen-free environment (nitrogen, sodium sulfite)and air-saturated environment | |
Storage stability | 24 months provided the sensor material is stored in the dark |