Small Trace Oxygen Sensor for Contactless Measurement

Oxygen Sensor Spot SP-PSt8-NAU

These oxygen sensors measure the partial pressure of both dissolved and gasseous oxygen. They are fixed to the inner surface of glass or transparent plastic containers and can then be read out from the outside – through the vessel wall. The trace oxygen sensors type PSt8 can be used for measurements in trace oxygen range of 0 – 10 % O2 and have a detection limit of 0.007 % O2. The SP-PSt8-NAU is not suitable for autoclaving.

  • Non-invasive measurements through the vessel wall
  • No consumption of oxygen
  • Signal independent of flow velocity
  • Measures oxygen in liquids as well as in gas phase
Get a quote Place an order

Applications

Bioprocess Development: Oxygen Monitoring in Shake Flasks

O2 supply is one of the major issues in the cultivation of aerobic organisms. Shake flask cultures are widely applied in academic and industrial bioprocess development. As adequate methods for real monitoring of dissolved oxygen were missing, sufficient O2 supply is usually assumed. The non-invasive oxygen sensors in shake flasks now ensure oxygen supply and give new insights into metabolic activity.


Respiration & Photosynthesis: Oxygen Monitoring in Glass Vials

Determination of respiratory activity is often performed for water organisms such as invertebrates, larval stages or eggs, but also for bacteria, cell cultures, yeasts or fungi. For algae measurement of photosynthetic activity is of great interest. Using our 20 mL SensorVial with an integrated sensor stripe oxygen can be measured simultaneously in the liquid sample and in the headspace. Autoclavable SensorVials for stirred and non-stirred applications are available.


Technical

Specifications Gaseous & Dissolved O2 Dissolved O2
*after two-point calibration as described in the manual
Measurement range 0 – 10 % O2
0 – 100 hPa

0 – 4.5 mg/L

0 – 140 µmol/L
Limit of detection 0.007 % oxygen 3 ppb
Resolution

± 0.002 % O2 at 0.008 % O2
± 0.06 % O2 at 2.5 % O2

± 0.7 ppb at 3 ppb
± 2.5 ppb at 1000 ppb

Accuracy at + 20 °C* ± 3 ppb or ± 3 % rel.
Measurement temperature range from 0 to + 50 °C
Response time (t90) < 6 sec. < 40 sec.
Properties    
Compatibility Aqueous solutions, ethanol, methanol
No cross-sensitivity

pH 1 – 14
CO2, H2S, SO2
Ionic species

Cross-sensitivity

Organic solvents, such as acetone, toluene, chloroform or methylene chloride
Chlorine gas

Sterilization procedure Ethylene oxide (EtO)
Gamma irradiation
Cleaning procedure

3 % H2O2
Ethanol
Aqueous solutions

Calibration Two-point calibration with oxygen-free environment (nitrogen, sodium sulfite) and a second calibration value optimally between 2 % and 4 % oxygen
Storage stability 60 months provided the sensor material is stored in the dark at room temperature

Related products

Microx 4 trace Stand-alone Fiber Optic Oxygen Meter for Contactless and Micro-invasive Measurements

Microx 4 trace

The Microx 4 trace is a completely stand-alone, portable…

More
OXY-1 ST trace Small & Versatile Trace Oxygen Meter for Use with Different Sensor Designs

OXY-1 ST trace

The OXY-1 ST trace is very compact and can be set up almost…

More
Polymer Optical Fiber POF with two SMA sockets

Polymer Optical Fiber POF

A polymer optical fiber (POF) is needed to transfer…

More
SOA Stick-On Adapter for Polymer Optical Fibers

Stick-On Adapter SOA

The stick-on adapter is used to attach the polymer optical…

More
Adapter for Round Containers with ST socket ARC1-ST

Adapter for Round Containers ARC1-ST

The Adapter for Round Containers ARC1-ST is an adjustable…

More
OXY-4 ST trace (G3) Versatile Multi-Channel Trace O2 Meter

OXY-4 ST trace (G3)

The OXY-4 ST trace can be used with O2 microsensor in…

More

Resources

Software

×

Presens TV

Watch tutorials, webinars and informative videos about PreSens optical sensor systems.

All videos