Small Trace Oxygen Sensor for Contactless Measurement
Oxygen Sensor Spot SP-PSt8-NAU
These oxygen sensors measure the partial pressure of both dissolved and gasseous oxygen. They are fixed to the inner surface of glass or transparent plastic containers and can then be read out from the outside – through the vessel wall. The trace oxygen sensors type PSt8 can be used for measurements in trace oxygen range of 0 – 10 % O2 and have a detection limit of 0.007 % O2. The SP-PSt8-NAU is not suitable for autoclaving.
- Non-invasive measurements through the vessel wall
- No consumption of oxygen
- Signal independent of flow velocity
- Measures oxygen in liquids as well as in gas phase
Applications
Bioprocess Development: Oxygen Monitoring in Shake Flasks
O2 supply is one of the major issues in the cultivation of aerobic organisms. Shake flask cultures are widely applied in academic and industrial bioprocess development. As adequate methods for real monitoring of dissolved oxygen were missing, sufficient O2 supply is usually assumed. The non-invasive oxygen sensors in shake flasks now ensure oxygen supply and give new insights into metabolic activity.
Respiration & Photosynthesis: Oxygen Monitoring in Glass Vials
Determination of respiratory activity is often performed for water organisms such as invertebrates, larval stages or eggs, but also for bacteria, cell cultures, yeasts or fungi. For algae measurement of photosynthetic activity is of great interest. Using our 20 mL SensorVial with an integrated sensor stripe oxygen can be measured simultaneously in the liquid sample and in the headspace. Autoclavable SensorVials for stirred and non-stirred applications are available.
Technical
Specifications | Gaseous & Dissolved O2 | Dissolved O2 |
---|---|---|
*after two-point calibration as described in the manual | ||
Measurement range | 0 – 10 % O2 0 – 100 hPa | 0 – 4.5 mg/L 0 – 140 µmol/L |
Limit of detection | 0.007 % oxygen | 3 ppb |
Resolution | ± 0.002 % O2 at 0.008 % O2 | ± 0.7 ppb at 3 ppb |
Accuracy at + 20 °C* | ± 3 ppb or ± 3 % rel. | |
Measurement temperature range | from 0 to + 50 °C | |
Response time (t90) | < 6 sec. | < 40 sec. |
Properties | ||
Compatibility | Aqueous solutions, ethanol, methanol | |
No cross-sensitivity | pH 1 – 14 | |
Cross-sensitivity | Organic solvents, such as acetone, toluene, chloroform or methylene chloride | |
Sterilization procedure | Ethylene oxide (EtO) Gamma irradiation | |
Cleaning procedure | 3 % H2O2 | |
Calibration | Two-point calibration with oxygen-free environment (nitrogen, sodium sulfite) and a second calibration value optimally between 2 % and 4 % oxygen | |
Storage stability | 60 months provided the sensor material is stored in the dark at room temperature |
Related products
Resources
FAQs
Manuals
Brochures
Software
PreSens Oxygen Calculator version 3.1.1 - Windows 7/8/10