Top

Measure Dissolved & Gaseous Oxygen Traces

Trace Range Oxygen Probe OXYPro® TR & TRM
(max. 0 - 10 % O2)

The OXYPro® TR & TRM combine an optical sensor type PSt6 and electro-optical module in one probe. The sensor has a maximum measurement range of 0 - 10 % O2, and is integrated in a removable sensor cap (OEC), which can easily be exchanged. These trace range probes are perfectly suited for application in the brewing industries. The sensor caps are not only available in a version safe for food applications but also for different other demands like fast response time, or for measurements in aggressive environments. The OXYPro® are directly connected to the control unit and have a standard PG 13.5 stainless steel thread for easy integration at the measurement site. As a digital interface these probes use RS485 (OXYPro® TR = PreSens proprietary, OXYPro® TRM = Modbus RTU) and have 4 - 20 mA analog output. A serial communication protocol is offered for data exchange between the OXYPro® and a PC or other host unit. For initial operation and adjustment of the 4 - 20 mA output the PreSens EOM-STS software can be used.

  • Maximum measurement range of 0 - 10 % O2, detection limit 0.002 % O2 / 1 ppb
  • Ideal for O2 measurements in breweries
  • Different versions: food applications, reduced bubble formation, aggressive environments
  • Stands SIP (recommended: + 121 °C, 30 min) & CIP (+ 80 °C, 2 % NaOH)
  • Simple integration due to standard PG 13.5 thread & Varivent adapter
  • Communication via RS485 (PreSens proprietary, Modbus RTU)
  • 4 - 20 mA output
Get a quote Place an order

Applications

Process Control in the Beverage Industries

Even low oxygen concentrations influence shelf-life and the taste of certain beverages. Due to the extraordinary low detection limits and outstanding accuracy of the OXYPro® MR and OXYPro® TR, the quality of oxygen-sensitive products such as beer, wine and soft drinks can be secured. With OXYPro® certain processes can be monitored and controlled. As the probes are autoclavable, stand SIP and CIP they are ideally suited for application in the beverage industries.


Monitoring Industrial Processes

The OXYPro® series probes are designed for harsh measurement conditions. Additionally they are almost maintenance-free as they do not contain any membranes and electrolytes, so they are ideally suited for monitoring chemical industrial processes or oxygen sensitive applications. They can be used for gaseous or dissolved oxygen monitoring in most different environments, like e. g. tubes or vessels.


Technical

Specifications Gaseous & Dissolved O2 Dissolved O2
*at 20 °C, 960 - 980 hPa; humidified gas mixture
**after customized calibration
***determined changing from 10 % oxygen gas to nitrogen 5.0
****determined changing from air-saturated water to a freshly prepared 1 % sodium sulfite solution
*****not for OEC-PSt6-NAU-OIW; this sensor cap cannot be sterilized or autoclaved
Measurement range* Optimal: 0 - 5 % O2, 0 - 50 hPa
Max.**: 0 - 10 % O2, 0 - 100 hPa
Optimal: 0 - 2 mg/L
Max.**: 0 - 4.5 mg/L
Limit of detection* 0.001 % O2, 0.01 hPa 0.5 ppb
Resolution* 0.002 % O2 ± 0.0006 % O2
0.2 % O2 ± 0.001 % O2
1 ppb ± 0.3 ppb
20 ppb ± 0.4 ppb
Response time (t90) < 10 sec.*** < 30 sec.****
Properties    
Compatibility Aqueous solutions
No cross-sensitivity to pH 1 - 14, CO2, SO2, ionic species
Cross-sensitivity to Organic solvents, such as pure acetone, toluene, chloroform or methylene chloride
Chlorine gas
Sterilization procedure***** Steam sterilization (recommended: + 121 °C, 30 min)
Cleaning procedure***** Cleaning in place (CIP, 2 % NaOH, + 80 °C, + 176  °F)
3 % H2O2, acidic agents (HCl, H2SO4) max. 4 - 5 %
Ethanol, methanol, cleaning agents at room temperature
Storage stability of sensor cap 5 years provided the sensor material is stored at room temperature in dry conditions and in the dark
Sampling rate 1 sec. up to 9 min. 59 sec.
Temperature sensor NTC
Accuracy of temperature measurement ± 1 °C in operation temperature range (0 °C to + 50 °C)
Power supply 7 - 24 V DC
Power consumption in active mode Max. 1 W
Power consumption in stand-by mode 0.15 W
Temperature range operation Optimal: 0 °C to + 50 °C
Max.: - 10 °C to + 90 °C
Storage conditions Optimal: room temperature (+ 20 °C ± 5  °C)
Max.: 0 °C to + 70 °C, 80 % rel. humidity
Pressure resistance at 20 °C - 1 bar to + 12 bar
Dimensions Max. immersion depth: 120 mm / 225 mm / 325 mm / 425 mm + 7 mm with OEC30
PG 13.5 thread and VP8 connector: 55 mm
Diameter: 12 mm
Probe housing material Stainless steel 1.4435
Surface quality of steel Ra < 0.4 μm (N5)
Wetted part Stainless steel 1.4435, silicone, EPDM
Weight 100 g
Digital interface RS485, half-duplex (variable Baud rate, default: 19200, data bits: 8, parity: none, stop bits: 1, handshake: none) or
RS485 Modbus RTU
Analog output 4 - 20 mA (current sink)

Related products

VP8-BT ST Cable with Bluetooth Module for OXYPro®

VP8-BT ST Cable with Bluetooth Module for OXYPro®

This accessory for OXYPro® allows controlling oxygen…

More
PreSens Wireless Studio Mobile App for Bluetooth-Capable PreSens Oxygen Measurement Devices

PreSens Wireless Studio

PreSens Wireless Studio is a mobile app that allows…

More
Oxygen Exchange Cap OEC-PSt6-NAU for OXYPro®

Oxygen Exchange Cap OEC-PSt6-NAU

The Oxygen Exchange Cap with integrated sensor type PSt6 is…

More
Oxygen Exchange Cap OEC-PSt6-YAU for OXYPro & OIMs

Oxygen Exchange Cap OEC-PSt6-YAU

The Oxygen Exchange Cap with integrated sensor type PSt6 is…

More
Oxygen Exchange Cap OEC30-PSt6-YAU for OIMs & OXYPro®

Oxygen Exchange Cap OEC30-PSt6-YAU

The Oxygen Exchange Cap OEC30 with integrated sensor type…

More
Oxygen Exchange Cap OEC30-PSt6-NAU

Oxygen Exchange Cap OEC30-PSt6-NAU

The Oxygen Exchange Cap OEC30 with integrated sensor type…

More
OXYPro® NPT Process Connection Adapter

OXYPro® NPT

This adapter has a NPT 1/2 thread and is made of PVDF. With…

More
OXYPro® FTM

OXYPro® FTM

The OXYPro® FTM is a metal flow-through connector that can…

More
OXYPro® Triclamp Adapter

OXYPro® Triclamp Adapter

The Triclamp Adapter is an accessory that allows to safely…

More
Adapter for 25 mm Ports OAD-D25

Adapter for 25 mm Ports OAD-D25

All PreSens OXYPro® series probes and Oxygen Probes for…

More
Handheld USB-QR Barcode Scanner

Handheld USB-QR Barcode Scanner

This handheld scanner can simply be connected to your PC via…

More
×

Presens TV

Watch tutorials, webinars and informative videos about PreSens optical sensor systems.

All videos