Easy and Fast Oxygen Measurements in Headspace
Needle-Type Oxygen Microsensor with Fixed Sensor Tip NFSG-PSt7
The NFSG-PSt7 is specially designed for headspace measurement in packaging or other applications where gaseous oxygen measurements in small compartments have to be conducted. The sensor tip of this microsensor is fixed inside a steel needle, so it does not protrude the needle tip. Secured this way inside the needle, the sensor can be pierced through material and the oxygen content in smallest headspace volumes can be determined. The NFSG-PSt7 has a measurement range of 0 – 50 % oxygen, with a limit of detection of 0.03 % oxygen.
- High spatial resolution (250 µm)
- High temporal resolution (t90 < 3 sec.)
- Penetration probe for insertion into packaging, like e.g. blister packaging
- Measurement in headspace
- No consumption of oxygen
- Signal independent of flow velocity
- Measurement in headspace or smallest volumes
Technical
| Specifications | Gaseous O2 | |
|---|---|---|
| *at + 20 °C, after two-point calibration **TF = flat sensor tip with 250 µm diameter | ||
| Measurement range | 0 – 50 % O2 0 – 5000 hPa | |
| Limit of detection | 0.03 % oxygen | |
| Resolution | ± 0.01 % O2 at 1 % O2 | |
| Accuracy at + 20 °C* | ± 0.05 % O2 or ± 3 % rel. | |
| Measurement temperature range | from 0 to + 50 °C | |
| Response time (t90) | TF**: < 3 sec. (gas) | |
| Properties | ||
| No cross-sensitivity | pH 1 – 14 | |
| Cross-sensitivity | Organic solvents, such as acetone, toluene, chloroform or methylene chloride | |
| Disinfection procedure | Disinfection with EtOH (70 % v/v) or isopropanol (100 % v/v) possible | |
| Cleaning procedure | Ethanol, soap solution, ddw | |
| Calibration | Two-point calibration with oxygen-free environment (nitrogen, sodium sulfite) and air-saturated environment | |