Cells for Online Oxygen Monitoring in Perfusion Systems
O2 Flow-Through Cell FTC-PSt7
A miniaturized oxygen sensor type PSt7 is integrated in a flow-through cell. It allows non-invasive online monitoring of oxygen in perfusion systems. The sensor is fixed to a color coded stick, which can be attached to flow-through cells of different size and shape according to requirements. A polymer optical fiber connects the sensor inside the flow-through cell to the respective O2 meter. The cells are made of polycarbonate.
- Precise online monitoring of oxygen
- Different sizes and shapes for various flow rates
- Easy connection to external tubing
Technical
Specifications | Dissolved O2 |
---|---|
*after two-point calibration as described in the manual **equilibrated FTC with physiological solution and sufficient flow rate (min. 15 mL/min) at 37 °C ***recalibration may be required | |
Measurement range | 0 – 45 mg/L 0 – 1400 µmol/L |
Resolution | ± 0.005 mg/L at 0.4mg/L ± 0.025 mg/L at 9.06 mg/L |
Accuracy * | at + 20 °C: ± 0.05 % O2 or ± 3 % rel. |
Measurement temperature range | From 0 to + 50 °C |
Response time (t90)** | < 40 sec. |
Properties | |
Compatibility | Aqueous solutions, ethanol, methanol |
Cross-sensitivity | Organic solvents, such as acetone, toluene, chloroform or methylene chloride Chlorine gas |
Sterilization procedure*** | Irradiation Ethylene oxide (EtO) |
Calibration | Sensor sticks are pre-calibrated; Two-point calibration in oxygen-free environment (nitrogen, sodium sulfite) and air-saturated environment |
Storage stability | Up to 60 months provided the sensor material is stored in the dark at room temperature |
T-cell formats | Luer T-cell (delivered), inner diameter 5 mm, cell volume 0.3 mL; 1/4" x 1/4" (Qosina), cell volume 2.1 mL; 3/8" x 3/8" (Qosina), cell volume 4.6 mL; 1/2" x 1/2" (Qosina), cell volume 8.3 mL |
Related products
Resources
FAQs
How can I convert an oxygen value into a different oxygen unit?
How can I prepare the calibration solutions cal0 and cal100 for oxygen sensors?
How do I import data into excel from a text file containing measurement data?
How do I verify whether the oxygen sensor is giving correct readings (performance proof)?
How does an oxygen sensor work?
How does salinity affect the oxygen measurement?
How does temperature affect the oxygen measurement?
The Stern-Volmer-equation
What is the maximum flow rate and pressure for your oxygen flow-through cells?
What is the time of delivery?
Which substances can interfere with the optical O2, pH and CO2 measurements?