Fast Oxygen Monitoring in Perfusion Systems

Flow-Through Cell Housed Oxygen Microsensor FTCH-PSt1

The oxygen microsensor housed in a flow-through cell is optimized for a fast response time (t90< 1 sec in gas, < 5 sec. in liquid). The microsensor with a tip size from < 50 µm to 140 µm is integrated in a sleeve made of Teflon® tightened with a silicone rubber, which is mounted to a T-connector. The liquid volume inside the FTCH is about 310 µL. External tubing with an inner diameter of 1.6, 2.3, and 3.2 mm can be connected via Luer-Lock adpaters. Liquids (e. g. water, blood, etc.) can be pumped through the cell. The FTCH-PSt1 has a measurement range of 0 – 50 % oxygen, with a detection limit of 0.05 % oxygen.

  • Perfusion monitoring in smallest volumes
  • High temporal resolution (< 5 sec.)
  • No consumption of oxygen
  • Measures in liquids as well as in the gas phase
Get a quote Place an order

Technical

Specifications Gaseous & Dissolved O2 Dissolved O2

*after two-point calibration as described in the manual
**TS: tapered sensor tip with a diameter < 50 μm and no optical isolation
***TF: flat-broken sensor tip with a diameter of 140 μm
 

Measurement range

0 – 50 % O2

0 – 500 hPa

0 – 22.5 mg/L

0 – 700 µmol/L
Limit of detection 0.05 % oxygen 20 ppb
Resolution

± 0.01 % O2 at 0.21 % O2
± 0.09 % O2 at 20.9 % O2
±0.1 hPa at 2 hPa

± 1 hPa at 207 hPa

± 0.005 mg/L at 0.09 mg/L
± 0.04 mg/L at 9.06 mg/L
± 0.14 µmol/L at 2.83 µmol/L

± 1.3 µmol/L at 283 µmol/L
Accuracy* ± 0.4 % O2 at 20.9 % O2
± 0.05 % O2 at 0.2 % O2
 
Drift < 0.1 % O2 within 30 days (sampling interval of 1 min.)
Measurement temperature range from 0 to + 50 °C
Response time TS** (t90) < 1 sec. (gas) < 2 sec. (liquid)
Response time TF*** (t90) < 15 sec. (gas) < 30 sec. (liquid)
Properties    
Compatibility Aqueous solutions, ethanol, methanol
No cross-sensitivity

pH 1 – 14
CO2, H2S, SO2
Ionic species

Cross-sensitivity Organic solvents, such as acetone, toluene, chloroform or methylene chloride
Chlorine gas
Sterilization procedure

Ethylene oxide (EtO)

Cleaning procedure 3 % H2O2
Ethanol
Soap solution
Calibration Two-point calibration with oxygen-free environment (nitrogen, sodium sulfite) and air-saturated environment
Storage stability 60 months provided the sensor material is stored in the dark at room temperature

Related products

Microx TX3 Micro Fiber Optic Oxygen Meter

Microx TX3

The Microx TX3 is a precise micro fiber optic oxygen meter.…

More
Microx TX3 trace Micro Fiber Optic Oxygen Meter

Microx TX3 trace

The Microx TX3 trace is a precise micro fiber optic oxygen…

More
Multi-channel fiber optic oxygen meter OXY-4 micro

OXY-4 micro

The OXY-4 micro is a precise multi-channel micro fiber optic…

More
Multi-channel fiber optic oxygen meter OXY-10 micro

OXY-10 micro

The OXY-10 micro is a precise multi-channel oxygen meter for…

More
×

Presens TV

Watch tutorials, webinars and informative videos about PreSens optical sensor systems.

All videos