Bare Fiber Oxygen Microsensor for Customized Measurement Set-Ups

Implantable Oxygen Microsensor IMP-PSt7

This optical oxygen microsensor can be applied in various customized applications. The tiny probe has a tip size < 50 µm to 230 µm while the outer diameter ranges from 230 µm to 900 µm. The microsensor tip is not mounted into any additional housing. The bare glass fiber tip can be mounted to your own housings, steel tubes and micro respirometer chambers etc. The small outer diameter even allows insertion into implantable Venflon-tubes. Moreover the microsensor can be deployed in sealed containers to measure the oxygen content directly. The IMP-PSt7 has a measurement range of 0 – 100 % oxygen, with a limit of detection of 0.03 % O2.

  • High spatial resolution (< 50 μm)
  • High temporal resolution (t90 < 3 sec.)
  • Integration into customized housings and set-ups
  • No consumption of oxygen
  • Measurement in headspace & liquids
Get a quote Place an order

Technical

Specifications* Gaseous & Dissolved O2 Dissolved O2
*data for microsensor TF = flat-broken sensor tip with 230 μm diameter
**after two-point calibration as described in the manual
Measurement range 0 – 100 % O2
0 – 1000 hPa

0 – 45 mg/L

0 – 1400 µmol/L
Limit of detection 0.03 % oxygen 15 ppb
Resolution

± 0.01 % O2 at 1  % O2
± 0.05 % O2 at 20.9 % O2

± 0.005 mg/L at 0.4 mg/L

± 0.025 mg/L at 9.06 mg/L
Accuracy at + 20 °C** ± 0.05 % O2 or ± 3 % rel.
Measurement temperature range from 0 to + 50 °C
Response time (t90) < 3 sec. (gas) < 10 sec. (liquid)
Properties    
Compatibility Aqueous solutions, ethanol, methanol
No cross-sensitivity

pH 1 – 14
CO2, H2S, SO2
Ionic species

Cross-sensitivity

Organic solvents, such as acetone, toluene, chloroform or methylene chloride
Chlorine gas

Sterilization procedure

Steam sterilization
Ethylene oxide (EtO)
Recalibration recommended

Cleaning procedure

3 % H2O2
Ethanol
Soap solution

Calibration Two-point calibration with oxygen-free environment (nitrogen, sodium sulfite) and air-saturated environment
Storage stability 60 months provided the sensor material is stored in the dark at room temperature

Related products

Microx 4 handheld optical O2 meter

Microx 4

The Microx 4 is a completely stand-alone, portable fiber…

More
Microx 4 trace Stand-alone Fiber Optic Oxygen Meter for Contactless and Micro-invasive Measurements

Microx 4 trace

The Microx 4 trace is a completely stand-alone, portable…

More
OXY-1 ST Compact Oxygen Transmitter for Use with Numerous Sensor Designs

OXY-1 ST

This extremely small and lightweight oxygen meter can be…

More
OXY-1 ST trace Small & Versatile Trace Oxygen Meter for Use with Different Sensor Designs

OXY-1 ST trace

The OXY-1 ST trace is very compact and can be set up almost…

More
Manual Micromanipulator with Needle-type O2 Microsensor

Manual Micromanipulator MM

The Manual Micromanipulator MM is specifically designed for…

More
MM33 Manual Micromanipulator

Manual Micromanipulator MM33

The Manual Micromanipulator MM33 offers high resolution…

More
Automated Micromanipulator with Heavy Stand

Automated Micromanipulator AM

The Automated Micromanipulator is specifically designed for…

More
PreSens Profiling Studio Software screenshots

PreSens Profiling Studio

This software enables control of the Automated…

More
OXY-4 ST (G3) Multi-Channel O2 Meter for Use with Numerous Sensor Designs

OXY-4 ST (G3)

Due to its small outer dimensions OXY-4 ST can be set up…

More
OXY-4 ST trace (G3) Versatile Multi-Channel Trace O2 Meter

OXY-4 ST trace (G3)

The OXY-4 ST trace can be used with O2 microsensor in…

More

Resources

Software

×

Presens TV

Watch tutorials, webinars and informative videos about PreSens optical sensor systems.

All videos