Top

Planar Trace Oxygen-Sensitive Spot

Oxygen Sensor Spot SP-PSt6-YAU

These sensor spots are the most versatile version of optical trace oxygen sensors. They can be attached to the inner surface of any transparent glass or plastic vessel like e. g. used for packaging. Oxygen is measured contactless and non-destructively through the transparent vessel wall. The SP-PSt6-YAU has a measurement range of 0 – 5 % oxygen in dissolved or gaseous phase. The oxygen sensitive coating is immobilized on a 1 mm glass support, which can be autoclaved.

  • Non-invasive measurements in a range of 0 - 5 % oxygen
  • No consumption of oxygen
  • Signal independent of flow velocity
  • Measures oxygen in liquids as well as in gas phase
  • Autoclavable
Get a quote Place an order

Applications

Bioprocess Development: Oxygen Monitoring in Shake Flasks

O2 supply is one of the major issues in the cultivation of aerobic organisms. Shake flask cultures are widely applied in academic and industrial bioprocess development. As adequate methods for real monitoring of dissolved oxygen were missing, sufficient O2 supply is usually assumed. The non-invasive oxygen sensors in shake flasks now ensure oxygen supply and give new insights into metabolic activity.


Respiration & Photosynthesis: Oxygen Monitoring in Glass Vials

Determination of respiratory activity is often performed for water organisms such as invertebrates, larval stages or eggs, but also for bacteria, cell cultures, yeasts or fungi. For algae measurement of photosynthetic activity is of great interest. Using our 20 mL SensorVial with an integrated sensor stripe oxygen can be measured simultaneously in the liquid sample and in the headspace. Autoclavable SensorVials for stirred and non-stirred applications are available.


Technical

Specifications Gaseous & Dissolved O2 Dissolved O2
*after two-point calibration as described in the manual
Measurement range

0 – 5  % O2

0 – 41.4 hPa

0 – 2 mg/L

0 – 56.9 µmol/L
Limit of detection 0.002 % oxygen

1 ppb

Resolution

± 0.0007 % O2 at 0.002 % O2
± 0.0015 % O2 at 0.2 % O2
± 0.007 hPa at 0.023 hPa

± 0.015 hPa at 2.0 hPa

± 0.0003 mg/L at 0.001 mg/L
± 0.0006 mg/L at 0.09 mg/L
± 0.010 µmol/L at  0.03µmol/L

± 0.020 µmol/L at 2.8 µmol/L
Accuracy* ± 1 ppb or ± 3 % of the respective concentration; whichever is higher
Drift at 0 % oxygen < 2 ppb within 30 days (sampling interval of 1 min.)
Measurement temperature range from 0 to + 50 °C
Response time (t90) < 6 sec. < 40 sec.
Properties    
Compatibility Aqueous solutions, ethanol, methanol
No cross-sensitivity

pH 1 – 14
CO2, H2S, SO2
Ionic species

Cross-sensitivity Organic solvents, such as acetone, toluene, chloroform or methylene chloride
Chlorine gas
Sterilization procedure

Steam sterilization
Ethylene oxide (EtO)

Cleaning procedure

Cleaning in place (CIP, 2 % NaOH, + 80 °C, + 176 °F)
3 % H2O2
Acidic agents (HCl, H2SO4), max. 4 % – 5 %

Calibration Two-point calibration with oxygen-free environment (nitrogen, sodium sulfite) and a second calibration value optimally between 1 % and 2 % oxygen
Storage stability 60 months provided the sensor material is stored in the dark

Related products

Fibox 4 trace handheld fiber optic O2 meter

Fibox 4 trace

The Fibox 4 trace is a stand-alone oxygen meter designed for…

More
OXY-1 SMA trace Small Optical Trace Oxygen Meter

OXY-1 SMA trace

As a successor of our established Fibox 3 trace, the OXY-1…

More
OXY-4 SMA trace (G3) Multi-Channel Trace Oxygen Meter

OXY-4 SMA trace (G3)

The OXY-4 SMA trace is compatible with sensor spots, dipping…

More
Fibox 3 trace Fiber Optic Trace Oxygen Meter

Fibox 3 trace

The Fibox 3 trace is a proven fiber optic device for precise…

More
Polymer Optical Fiber POF with two SMA sockets

Polymer Optical Fiber POF

A polymer optical fiber (POF) is needed to transfer…

More
SOA Stick-On Adapter for Polymer Optical Fibers

Stick-On Adapter SOA

The stick-on adapter is used to attach the polymer optical…

More
Adjustable Velcro®-type Adapter for Round Containers (ARC) with SMA sockets

Adapter for Round Containers ARC

The Adapter for Round Containers is an adjustable…

More
Coaster for Shake Flasks CFG with Polymer Optical Fiber for optical sensor read-out

Coaster for Shake Flasks CFG

The coaster CFG is placed underneath the shake flask with…

More

Resources

FAQs

Software

×

Presens TV

Watch tutorials, webinars and informative videos about PreSens optical sensor systems.

All videos